Drug loss from Paclitaxel-Coated Balloons During Preparation, Insertion and Inflation for Angioplasty: A Laboratory Investigation.


ABSTRACT:

Purpose

To investigate drug contamination of the working environment with paclitaxel drug-coated balloon (DCB) angioplasty due to loss of paclitaxel containing particles from the coating during DCB preparation, insertion, and inflation.

Material and methods

In an experimetal laboratory setting, drug loss during removal of the protective cover and insertion of the DCB through the hemostatic valve of the introducer sheath and after inflation was examined. In seven DCB types of different manufacturers, semi-quantitative image analysis was performed during five standardized tests cycles. Additionally, every DCB type passed one cycle of a wipe test and one cycle of air sampling.

Results

By removing the protective cover, the paclitaxel-covered balloon surface was significantly reduced in 3 out of 7 products (P = 0.043). Overall, extend of decline ranged from 0.4 to 12%. In 6 of 7 products, powdered paclitaxel clusters dropped down upon removal of the protective cover (0.099 ng/cm2 up to approx. 22 ng/cm2). Contamination of the air was detected in none of the DCB types. When pushed through the vascular sheath, none of the investigated DCB types showed a significant loss of paclitaxel from the coated balloon surface. After balloon inflation, the paclitaxel-coated surface area varied between manufacturers ranging from 25.9 to 97.8%.

Conclusion

In some DCB types, the removal of the protective cover already leads to a significant loss of paclitaxel and paclitaxel-coated surfaces. As a result, there will be a contamination of the workplace and a reduction in the therapeutic dose.

Level of evidence

No level of evidence.

SUBMITTER: Faenger B 

PROVIDER: S-EPMC9307540 | biostudies-literature | 2022 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Drug loss from Paclitaxel-Coated Balloons During Preparation, Insertion and Inflation for Angioplasty: A Laboratory Investigation.

Faenger Bernd B   Heinrich Andreas A   Hilger Ingrid I   Teichgräber Ulf U  

Cardiovascular and interventional radiology 20220610 8


<h4>Purpose</h4>To investigate drug contamination of the working environment with paclitaxel drug-coated balloon (DCB) angioplasty due to loss of paclitaxel containing particles from the coating during DCB preparation, insertion, and inflation.<h4>Material and methods</h4>In an experimetal laboratory setting, drug loss during removal of the protective cover and insertion of the DCB through the hemostatic valve of the introducer sheath and after inflation was examined. In seven DCB types of diffe  ...[more]

Similar Datasets

| S-EPMC9896391 | biostudies-literature
2023-03-02 | GSE217550 | GEO
| S-EPMC11858837 | biostudies-literature
| S-EPMC11617568 | biostudies-literature
| S-EPMC7560869 | biostudies-literature
| S-EPMC5226992 | biostudies-literature
| S-EPMC10632265 | biostudies-literature
| S-EPMC8396966 | biostudies-literature
| S-EPMC9809102 | biostudies-literature