ABSTRACT: Purpose
The purpose of this study was to identify the cause of severe image artifacts that occurred during gantry rotation in a 0.35 T MRI-Linac by comparing measurements of eddy currents, center frequency, and field inhomogeneities made with the gantry in motion and stationary.Methods
Gradient and B0 eddy currents were calculated from the free induction decays (FIDs) resulting from selective excitation at a temporal resolution of 200 ms/measurement. B0 eddy currents were also calculated from FIDs acquired with nonselective excitation at a temporal resolution of 100 ms/measurement. Center frequencies and B0 inhomogeneities were measured by acquiring FIDs with a repetition time (TR) of 290 ms. Cartesian and radial 2D true fast imaging with st