ABSTRACT: Introduction
Lupus anticoagulant (LA) testing is commonly performed within hemostasis laboratories, and the ACL TOP 50 family of instruments represent a new "single platform" of hemostasis instrumentation. Our aim was to evaluate these instruments and manufacturer reagents or alternatives for utility in LA testing.Methods
Comparative evaluations of LA testing using newly installed ACL TOPs 550 and 750 as well as comparative assessments with existing "reference," predominantly Stago, instrumentation, and reagents. Evaluations comprised both dilute Russell viper venom time (dRVVT) and activated partial thromboplastin time (APTT)-based assays. Establishment of normal reference ranges (NRR).Results
The HemosIL dRVVT-based assays showed good comparability with the existi