A multi-laboratory assessment of lupus anticoagulant assays performed on the ACL TOP 50 family for harmonized testing in a large laboratory network.


ABSTRACT:

Introduction

Lupus anticoagulant (LA) testing is commonly performed within hemostasis laboratories, and the ACL TOP 50 family of instruments represent a new "single platform" of hemostasis instrumentation. Our aim was to evaluate these instruments and manufacturer reagents or alternatives for utility in LA testing.

Methods

Comparative evaluations of LA testing using newly installed ACL TOPs 550 and 750 as well as comparative assessments with existing "reference," predominantly Stago, instrumentation, and reagents. Evaluations comprised both dilute Russell viper venom time (dRVVT) and activated partial thromboplastin time (APTT)-based assays. Establishment of normal reference ranges (NRR).

Results

The HemosIL dRVVT-based assays showed good comparability with the existing Stago reference method (R > 0.9) and could be considered as verified as fit for purpose. A variety of APTT assays was additionally evaluated for LA utility, and we identified from the assessment good utility of a non-Werfen solution in Hyphen BioMed Cephen reagents. NRR were established based on ≥120 normal individual plasma samples.

Conclusion

This evaluation of LA reagents on ACL TOP 50 Family instruments identified overall acceptable performance of both dRVVT (Werfen solution) and APTT (non-Werfen solution) to enable harmonization of LA testing in our large network.

SUBMITTER: Favaloro EJ 

PROVIDER: S-EPMC9311435 | biostudies-literature | 2022 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

A multi-laboratory assessment of lupus anticoagulant assays performed on the ACL TOP 50 family for harmonized testing in a large laboratory network.

Favaloro Emmanuel J EJ   Mohammed Soma S   Vong Ronny R   Chapman Kent K   Swanepoel Priscilla P   Kershaw Geoffrey G   Cai Nancy N   Just Sarah S   Connelly Lynne L   Brighton Timothy T   Pasalic Leonardo L  

International journal of laboratory hematology 20220301 3


<h4>Introduction</h4>Lupus anticoagulant (LA) testing is commonly performed within hemostasis laboratories, and the ACL TOP 50 family of instruments represent a new "single platform" of hemostasis instrumentation. Our aim was to evaluate these instruments and manufacturer reagents or alternatives for utility in LA testing.<h4>Methods</h4>Comparative evaluations of LA testing using newly installed ACL TOPs 550 and 750 as well as comparative assessments with existing "reference," predominantly Sta  ...[more]

Similar Datasets

| S-EPMC10278396 | biostudies-literature
| S-EPMC8922544 | biostudies-literature
| S-EPMC7938630 | biostudies-literature
| S-EPMC9601296 | biostudies-literature
| S-EPMC8935533 | biostudies-literature
| S-EPMC4127163 | biostudies-literature
| S-EPMC8669994 | biostudies-literature
| S-EPMC7376154 | biostudies-literature
| S-EPMC8200390 | biostudies-literature