An evaluation of M2+ interference correction approaches associated with As and Se in ICP-MS using a multi-day dataset along with ICP-MS/MS/HR-ICP-MS based analysis and hierarchical modeling as a means of assessing bias in fortified drinking waters and single component matrices.
Ontology highlight
ABSTRACT: Three 12 mass oriented rare earth element (REE) M2+ correction approaches (fixed factor, a dual internal standard, and an in-sample) are evaluated for use in an ICP-MS environmental method update. The multi-variant-based evaluation includes analyzing the same 19 REE-fortified matrices on eight different days over a two-month period using two instrument tunes. These REE-fortified matrices were also analyzed using HR-ICP-MS and ICP-MS/MS to estimate the reference value for use in the principal component analysis (PCA) and hierarchical modeling evaluation. A fixed factor is unable to compensate for matrix and mass dependent drift and because of this it generates the largest across matrix, tune, and day 95th percent confidence bounds for the REE corrections on both As (1.