Unknown

Dataset Information

0

Gap distribution mapping to visualize regions associated with type 1 endoleak in a fenestrated thoracic stent graft.


ABSTRACT:

Objectives

Our goal was to analyse the relationships between aortic surface irregularity and a type 1 endoleak (T1EL) after a thoracic endovascular repair using the Najuta fenestrated stent graft.

Methods

The patients who were treated using the Najuta stent graft for an intact aortic arch aneurysm at the Saitama Cardiovascular and Respiratory Center between June 2013 and June 2017 were analysed retrospectively. The primary end point was the occurrence of a T1EL. The gap between a virtual aorta and the patient's aortic wall at the stent graft fixation was calculated over the whole circumference at 1 mm intervals, and gap distribution mapping was performed. The rate of freedom from a T1EL was estimated using the Kaplan-Meier method and compared between the patients with or without a continuously malapposed region of >1 mm from the branches to the aneurysm.

Results

Twenty-one patients were analysed. During the mean follow-up period of 21.7 months, 4 patients were confirmed to have T1ELs. Three of the T1Els were detected during the perioperative period and occurred through a fenestration. The remaining patient had a T1EL 2 years postoperatively. The gap distribution mapping confirmed the presence of a continuously malapposed region of >1 mm from the cervical branch to the aneurysm postoperatively in 4 patients with T1ELs. Patients who had a continuously malapposed region of >1 mm showed a statistically lower T1EL rate than those without (p < 0.001). Malapposed regions defined using the gap distribution mapping were consistent with flow channels through T1EL fenestrations detected using the perioperative computed tomography data.

Conclusions

The gap distribution mapping could be useful to predict the occurrence of T1ELs in patients with the Najuta stent graft.

SUBMITTER: Shukuzawa K 

PROVIDER: S-EPMC9319607 | biostudies-literature | 2022 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Gap distribution mapping to visualize regions associated with type 1 endoleak in a fenestrated thoracic stent graft.

Shukuzawa Kota K   Fujii Tomoya T   Sumi Makoto M   Kozaki Junya J   Umezu Mitsuo M   Ohki Takao T   Iwasaki Kiyotaka K  

European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery 20220701 2


<h4>Objectives</h4>Our goal was to analyse the relationships between aortic surface irregularity and a type 1 endoleak (T1EL) after a thoracic endovascular repair using the Najuta fenestrated stent graft.<h4>Methods</h4>The patients who were treated using the Najuta stent graft for an intact aortic arch aneurysm at the Saitama Cardiovascular and Respiratory Center between June 2013 and June 2017 were analysed retrospectively. The primary end point was the occurrence of a T1EL. The gap between a  ...[more]

Similar Datasets

| S-EPMC9424361 | biostudies-literature
| S-EPMC9018899 | biostudies-literature
| S-EPMC8515168 | biostudies-literature
| S-EPMC11292094 | biostudies-literature
| S-EPMC8302879 | biostudies-literature
| S-EPMC10228724 | biostudies-literature
| S-EPMC11477123 | biostudies-literature
| S-EPMC7676711 | biostudies-literature
| S-EPMC10457440 | biostudies-literature
| S-EPMC7371721 | biostudies-literature