Free-breathing self-gated continuous-IR spiral T1 mapping: Comparison of dual flip-angle and Bloch-Siegert B1-corrected techniques.


ABSTRACT:

Purpose

To develop a B1-corrrected single flip-angle continuous acquisition strategy with free-breathing and cardiac self-gating for spiral T1 mapping, and compare it to a previous dual flip-angle technique.

Methods

Data were continuously acquired using a spiral-out trajectory, rotated by the golden angle in time. During the first 2 s, off-resonance Fermi RF pulses were applied to generate a Bloch-Siegert shift B1 map, and the subsequent data were acquired with an inversion RF pulse applied every 4 s to create a T1* map. The final T1 map was generated from the B1 and the T1* maps by using a look-up table that accounted for slice profile effects, yielding more accurate T1 values. T1 values were compared to those from inversion recovery (IR) spin echo (phantom only), MOLLI, SA