ABSTRACT: Objectives
3D chemical shift-encoded (CSE) MRI enables accurate and precise quantification of proton density fat fraction (PDFF) and R2*, biomarkers of hepatic fat and iron deposition. Unfortunately, 3D CSE-MRI requires reliable breath-holding. Free-breathing 2D CSE-MRI with sequential radiofrequency excitation is a motion-robust alternative but suffers from low signal-to-noise-ratio (SNR). To overcome this limitation, this work explores the combination of flip angle-modulated (FAM) 2D CSE imaging with a non-local means (NLM) motion-corrected averaging technique.Methods
In this prospective study, 35 healthy subjects (27 children/8 adults) were imaged on a 3T MRI-system. Multi-echo 3D CSE ("3D") and 2D CSE FAM ("FAM") images were acquired during breath-hold and free-breathin