ABSTRACT: Background
Critical care ventilators are frequently used to provide noninvasive ventilation (NIV) support to critically ill patients. Questions remain regarding carbon dioxide (CO2) clearance while using a critical care ventilator and dual limb circuit with various patient interfaces. The purpose of this study is to determine the positive end expiratory pressure (PEEP) level required to effectively washout CO2 for full-face and oronasal masks when using a dual limb circuit.Method
This randomized crossover trial was conducted at an academic medical center in the Midwest United States. After obtaining informed consent, eight healthy volunteers were placed on a 980 Puritan Bennett (Medtronic, Minneapolis, MN) ventilator operating in the NIV mode. All subj