Comparison of two different contrast sensitivity devices in young adults with normal visual acuity with or without refractive surgery.


ABSTRACT: This study investigated the reliability and correlation of two contrast sensitivity test (CST) devices in young adults with normal visual acuity, with or without refractive surgery. 57 patients aged 20-39 years who received both manual (OPTEC-6500) and automated CST (CGT-2000) examinations from June 19 to July 24, 2021 were retrospectively enrolled. Patients with corrected visual acuity under 20/20 or history of ocular surgery other than refractive surgery were excluded. 82 eyes of 41 patients (40 eyes with and 42 without history of refractive surgery) were enrolled. Mean time taken to complete each examination was 396.4 ± 20.4 and 286.8 ± 2.3 s using manual and automated CST, respectively (P < 0.001). Patients who underwent refractive surgery had significantly decreased area under the log

SUBMITTER: Jung H 

PROVIDER: S-EPMC9334259 | biostudies-literature | 2022 Jul