Detection of SARS-CoV-2 in different human biofluids using the loop-mediated isothermal amplification assay: A prospective diagnostic study in Fortaleza, Brazil.
Ontology highlight
ABSTRACT: We adopted the reverse-transcriptase-loop-mediated isothermal amplification (RT-LAMP) to detect severe acute respiratory syndrome coronavirus 2 (SARS-Cov-2) in patient samples. Two primer sets for genes N and Orf1ab were designed to detect SARS-CoV-2, and one primer set was designed to detect the human gene Actin. We collected prospective 138 nasopharyngeal swabs, 70 oropharyngeal swabs, 69 salivae, and 68 mouth saline wash samples from patients suspected to have severe acute respiratory syndrome (SARS) caused by SARS-CoV-2 to test the RT-LAMP in comparison with the gold standard technique reverse-transcription quantitative polymerase chain reaction (RT-qPCR). The accuracy of diagnosis using both primers, N5 and Orf9, was evaluated. Sensitivity and specificity for diagnosis were 96% (95%