Unknown

Dataset Information

Setup uncertainties and appropriate setup margins in the head-tilted supine position of whole-brain radiotherapy (WBRT).


ABSTRACT: Various applications of head-tilting techniques in whole-brain radiotherapy (WBRT) have been introduced. However, a study on the setup uncertainties and margins in head-tilting techniques has not been reported. This study evaluated the setup uncertainties and determined the appropriate planning target volume (PTV) margins for patients treated in the head-tilted supine (ht-SP) and conventional supine position (c-SP) in WBRT. Thirty patients who received WBRT at our institution between October 2020 and May 2021 in the c-SP and ht-SP were investigated. The DUON head mask (60124, Orfit Industries, Wijnegem, Belgium) was used in the c-SP, and a thermoplastic U-Frame Mask (R420U, Klarity Medical & Equipment Co. Ltd., Lan Yu, China) was used in the ht-SP. Daily setup verification using planning c

SUBMITTER: Park JW 

PROVIDER: S-EPMC9352041 | biostudies-literature | 2022