Unknown

Dataset Information

Ex vivo evaluation of personal protective equipment in hands-on defibrillation.


ABSTRACT:

Background

Defibrillation guidelines recommend avoiding patient contact during shock delivery. However, hands-on defibrillation (compressions during shock) and manual pressure augmentation (MPA - pushing on the defibrillator pads during shock) may lead to improved clinical outcomes. There are limited data addressing the protection provided by personal protective equipment (PPE) during hands-on defibrillation and MPA. This study investigated the hand-to-hand and hand-to-knee leakage current experienced by a simulated kneeling provider wearing different PPE.

Methods

A defibrillator was used in experiments on a pork shoulder, investigating three different hands-on positions: closed fist on defibrillator pads; open palm on pads with inadvertent finger contact (overhang); and ope

SUBMITTER: Stephens AF 

PROVIDER: S-EPMC9356271 | biostudies-literature | 2022 Sep

REPOSITORIES: biostudies-literature