Accuracy of Consumer-marketed smartphone-paired alcohol breath testing devices: A laboratory validation study.


ABSTRACT:

Background

Although alcohol breath testing devices that pair with smartphones are promoted for the prevention of alcohol-impaired driving, their accuracy has not been established.

Methods

In a within-subjects laboratory study, we administered weight-based doses of ethanol to two groups of 10 healthy, moderate drinkers aiming to achieve a target peak blood alcohol concentration (BAC) of 0.10%. We obtained a peak phlebotomy BAC and measured breath alcohol concentration (BrAC) with a police-grade device (Intoxilyzer 240) and two randomly ordered series of 3 consumer smartphone-paired devices (6 total devices) with measurements every 20 min until the BrAC reached <0.02% on the police device. Ten participants tested the first 3 devices, and the other 10 participants tested the ot