Unknown

Dataset Information

Safety and risk factors of TINAVI robot-assisted percutaneous pedicle screw placement in spinal surgery.


ABSTRACT:

Objective

To determine the rates and risk factors of pedicle screw placement accuracy and the proximal facet joint violation (FJV) using TINAVI robot-assisted technique.

Methods

Patients with thoracolumbar fractures or degenerative diseases were retrospectively recruited from June 2018 and June 2020. The pedicle penetration and proximal FJV were compared in different instrumental levels to identify the safe and risk segments during insertion. Moreover, the factors were also assessed using univariate and multivariate analyses.

Results

A total of 72 patients with 332 pedicle screws were included in the current study. The optimal and clinically acceptable screw positions were 85.8% and 93.4%. Of the 332 screws concerning the intra-pedicular accuracy, 285 screws (85.8%) w

SUBMITTER: Zhang RJ 

PROVIDER: S-EPMC9361479 | biostudies-literature | 2022 Aug

REPOSITORIES: biostudies-literature