Unknown

Dataset Information

Real-Time Risk Score for Glaucoma Mass Screening by Spectral Domain Optical Coherence Tomography: Development and Validation.


ABSTRACT:

Purpose

To develop and validate a risk score assessable in real-time using only retinal thickness-related values measured by spectral domain optical coherence tomography alone for use in population-based glaucoma mass screenings.

Methods

A total of 7572 participants (aged 35-74 years) underwent spectral domain optical coherence tomography examination annually between 2016 to 2021 in a population-based setting. We selected 284 glaucoma cases and 284 controls, matched by age and sex, from 11,487 scans in 2016. We conducted multivariable logistic regression with backward stepwise selection of retinal thickness-related variables to develop the diagnostic models. The developed risk scores were applied to all participants in 2018 (9720 eyes), and we randomly selected 723 scans for

SUBMITTER: Fukai K 

PROVIDER: S-EPMC9366724 | biostudies-literature | 2022 Aug

REPOSITORIES: biostudies-literature