Dosimetric Effects of Differences in Multi-Leaf Collimator Speed on SBRT-VMAT for Central Lung Cancer Patients.
Ontology highlight
ABSTRACT: Purpose: We aimed to investigate the effects of different multi-leaf collimator (MLC) speed constraints in volumetric modulated radiotherapy (VMAT) on the robustness of treatment plans for central lung cancer patients. Method and Materials: Twenty patients with central lung tumor who underwent stereotactic body radiotherapy (SBRT) with the VMAT technique at our hospital were included in this retrospective study. The reference plans were created with 3 different MLC speed constraints (Plan A: 0.1 cm/deg., Plan B: 0.3 cm/deg., and Plan C: 0.5 cm/deg.) with a 50-Gy/8Fr, planning target volume (PTV) D95% prescription. In each of these plans, setup errors from 1 to 5 mm were intentionally added in the direction of the central organ at 1-mm intervals (300 plans [20 cases
SUBMITTER: Saito M
PROVIDER: S-EPMC9379802 | biostudies-literature | 2022 Jan-Dec
REPOSITORIES: biostudies-literature
ACCESS DATA