Cross-calibration of two dual-energy X-ray absorptiometry devices for the measurement of body composition in young children.
Ontology highlight
ABSTRACT: This study aimed to cross-calibrate body composition measures from the GE Lunar Prodigy and GE Lunar iDXA in a cohort of young children. 28 children (mean age 3.4 years) were measured on the iDXA followed by the Prodigy. Prodigy scans were subsequently reanalysed using enCORE v17 enhanced analysis ("Prodigy enhanced"). Body composition parameters were compared across three evaluation methods (Prodigy, Prodigy enhanced, iDXA), and adjustment equations were developed. There were differences in the three evaluation methods for all body composition parameters. Body fat percentage (%BF) from the iDXA was approximately 1.5-fold greater than the Prodigy, whereas bone mineral density (BMD) was approximately 20% lower. Reanalysis of Prodigy scans with enhanced software attenuated these differences
SUBMITTER: Lyons-Reid J
PROVIDER: S-EPMC9381538 | biostudies-literature | 2022 Aug
REPOSITORIES: biostudies-literature
ACCESS DATA