Manufacturer Signal-to-Cutoff Threshold Underestimates Cumulative Incidence of SARS-CoV-2 Infection: Evidence from the Los Angeles Firefighters Study.


ABSTRACT:

Background

The objective of this analysis was to compare the performance sensitivity and specificity of manufacturer-recommended signal-to-cutoff (S/Co) thresholds with modified S/Co values to estimate the prevalence of SARS-CoV-2-specific antibodies in a cohort of firefighters with a known infection history.

Methods

Plasma venipuncture samples were used for serologic analysis of firefighters in Los Angeles, CA, USA, in October 2020. Seropositivity was assessed using the manufacturer's recommended S/Co (≥1.4 IgG) and modified S/Co thresholds based on measured antibody levels in 178 negative control patients who had blood drawn prior to the emergence of COVID-19. Optimal S/Co threshold was determined by receiver operating characteristic (ROC) curve analysis.

Results

Of 585 firefighters included in the study, 52 (8.9%) reported having a PCR-positive test history prior to antibody testing. Thirty-five (67.3%) firefighters with a previous PCR-positive test were seropositive based on the manufacturer S/Co thresholds, consistent with an estimated 67.3% sensitivity and 100% specificity. After evaluating multiple modified S/Co thresholds based on pre-pandemic negative samples, a modified S/Co of 0.36 was found to yield optimal sensitivity (88.5%) and specificity (99.4%) by ROC curve analysis. This modified threshold improved serostatus classification accuracy by 21.2%.

Conclusions

S/Co thresholds based on known negative samples significantly increase seropositivity and more accurately estimate cumulative incidence of disease compared to manufacturer-based thresholds.

SUBMITTER: Toubat O 

PROVIDER: S-EPMC9384175 | biostudies-literature | 2022 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Manufacturer Signal-to-Cutoff Threshold Underestimates Cumulative Incidence of SARS-CoV-2 Infection: Evidence from the Los Angeles Firefighters Study.

Toubat Omar O   Berg Anders H AH   Sobhani Kimia K   Mulligan Karen K   Hori Acacia M AM   Bhattacharya Jay J   Sood Neeraj N  

The journal of applied laboratory medicine 20220901 5


<h4>Background</h4>The objective of this analysis was to compare the performance sensitivity and specificity of manufacturer-recommended signal-to-cutoff (S/Co) thresholds with modified S/Co values to estimate the prevalence of SARS-CoV-2-specific antibodies in a cohort of firefighters with a known infection history.<h4>Methods</h4>Plasma venipuncture samples were used for serologic analysis of firefighters in Los Angeles, CA, USA, in October 2020. Seropositivity was assessed using the manufactu  ...[more]

Similar Datasets

| S-EPMC2867354 | biostudies-literature
| S-EPMC6298681 | biostudies-literature
| S-EPMC2905589 | biostudies-literature
| S-EPMC10922010 | biostudies-literature
| S-EPMC7946301 | biostudies-literature
| S-EPMC3311208 | biostudies-literature
| S-EPMC4156901 | biostudies-literature
| S-EPMC8067666 | biostudies-literature
| S-EPMC3621187 | biostudies-literature
| S-EPMC3309692 | biostudies-literature