ABSTRACT: Background
Dead-time correction is required for accurate quantitative SPECT-based dosimetry in the context of personalised 177Lu radiopharmaceutical therapy. We aimed to evaluate the impact of applying dead-time correction on the reconstructed SPECT image versus on the acquisition projections before reconstruction.Methods
Data from 16 SPECT/CT acquisitions of a decaying 177Lu-filled phantom (up to 20.75 GBq) and dual-timepoint SPECT/CT in 14 patients treated with personalised 177Lu peptide receptor radionuclide therapy were analysed. Dead time was determined based on the acquisition wide-spectrum count rate for each projection and averaged for the entire acquisition. Three dead-time correction methods (DTCMs) were used: the per-projection co