Agreement between Two Swept-Source Optical Coherence Tomography Biometers and a Partial Coherence Interferometer.


ABSTRACT:

Purpose

To evaluate the level of agreement between ANTERION (Heidelberg Engineering, Heidelberg, Germany), OA-2000 (Tomey, Nagoya, Japan), and IOLMaster 500 (Carl Zeiss AG, Jena, Germany).

Methods

Fifty-one eyes of 51 patients were included in the study. Flat keratometry (K) and steep K, vector component of astigmatism (Jackson cross-cylinder at 0° and 90° [J0] and Jackson cross-cylinder at 45° and 135° [J45]), anterior chamber depth, and axial length were compared using the three devices. Repeated measures analysis of variance was conducted to compare the mean values of the biometrics. Pearson correlation test was conducted to analyze the correlations of the measured values, and a Bland-Altman plot was used to assess the agreement between the three devices. The predicted in

SUBMITTER: Moon JY 