Unknown

Dataset Information

Preclinical validation and treatment of volumetric modulated arc therapy based total bone marrow irradiation in Halcyon™ ring gantry linear accelerator.


ABSTRACT:

Aim

This study aims​ to report preclinical validation, and the first clinical treatment of total bone marrow irradiation (TMI) and total bone marrow and lymph nodal irradiation (TMLI) using Volumetric modulated arc therapy in Halcyon-E ring gantry linear accelerator. Preclinical validation includes simulation, planning, patient-specific QA, and dry run.

Material and method

Four patients, two female and two male, with body weights of 116 kg, 52 kg, 64 kg, and 62 kg; with two with chronic myeloid leukemia, one each with acute lymphoblastic leukemia and acute myeloid leukemia (AML) were simulated and planned for TMI/TMLI. Patients were immobilized with a full-body vacuum bag. Head first supine (HFS) and Feet first supine (FFS) CT scans were acquired from head to knee and knee t

SUBMITTER: Shahid T 

PROVIDER: S-EPMC9389791 | biostudies-literature | 2022 Aug

REPOSITORIES: biostudies-literature