Ontology highlight
ABSTRACT: Purpose
In vivo dosimetry is a quality assurance tool that provides post-treatment measurement of the absorbed dose as delivered to the patient. This dosimetry compares the prescribed and measured dose delivered to the target volume. In this study, a tissue-equivalent water phantom provided the simulation of the human environment. The skin and entrance doses were measured using GafChromic EBT2 film for a Theratron® Equinox Cobalt-60 teletherapy machine.Methods
We examined the behaviors of unencapsulated films and custom-made film encapsulation. Films were cut to 1 cm × 1 cm, calibrated, and used to assess skin dose depositions and entrance dose. We examined the response of the film for variations in field size, source to skin distance (SSD), gantry angle and wedge angle.
SUBMITTER: Addo DA
PROVIDER: S-EPMC9390906 | biostudies-literature | 2022
REPOSITORIES: biostudies-literature