Calculating Relative Correction Factors for Quantitative Analysis with HILIC-HPLC-ELSD Method: Eight Fructooligosaccharides of Morinda Officinalis as a Case Study.


ABSTRACT:

Objective

Because the response of evaporating light scattering detector (ELSD) being in a nonlinear mode, there is no consensus on the method of calculating its relative correction factors (RCF), which limits the application of the quantitative analysis for multi-components by a single marker (QAMS) with LC-ELSD.

Methods

Using eight fructooligosaccharides of Morinda officinalis as a case study, the nystose (GF3) as a single standard was adopted to develop a QAMS method to simultaneously determine the other seven fructooligosaccharides with HILIC-HPLC-ELSD method. Six calculation methods of RCF were investigated to select the most reasonable method. The relative error of content between the QAMS and the external standard method (ESM) obtained from 30 batches of samples

SUBMITTER: Zhou L 

PROVIDER: S-EPMC9391178 | biostudies-literature | 2022

REPOSITORIES: biostudies-literature