Unknown

Dataset Information

Electrochemical Detection of Dihydroxybenzene Isomers at a Pencil Graphite Based Electrode.


ABSTRACT: In this work, an HB pencil electrode (HBPE) was electrochemically modified by amino acids (AAs) glycine (GLY) and aspartic acid (ASA) and designated as GLY-HB and ASA-HB electrodes. They were used in the detection of dihydroxybenzene isomers (DHBIs) such as hydroquinone (HQ), catechol (CC), and resorcinol (RS), by cyclic voltammetry (CV), and by differential pulse voltammetry. HBPE was characterized by scanning electron microscopy and energy-dispersive X-ray spectroscopy. These three electrodes showed a linear relationship of current with concentration of DHBIs, and the electrochemical processes were diffusion controlled in all cases. In simultaneous detection, the limit of detection, based on signal-to-noise ratio (S/N = 3), for HQ, CC, and RS was 12.473, 16.132, and 25.25 μM, respectivel

SUBMITTER: Khan MMI 

PROVIDER: S-EPMC9404491 | biostudies-literature | 2022 Aug

REPOSITORIES: biostudies-literature