Time-effectiveness and convenience of transvaginal ultrasound probe disinfection using ultraviolet vs chlorine dioxide multistep wipe system: prospective survey study.


ABSTRACT:

Objectives

To compare the efficiency, ease of use and user satisfaction of two methods of transvaginal ultrasound probe high-level disinfection: ultraviolet-C radiation (UV-C) and a chlorine dioxide multistep wipe system.

Methods

This was a prospective survey study. UV-C units were introduced into a busy early pregnancy assessment service and compared with a multiwipe system for disinfection. Before seeing each patient, healthcare professionals (HCPs) measured with a stopwatch the time taken to complete a cycle of disinfection using either UV-C or chlorine dioxide multistep wipes and responded to a quick-response (QR) code-linked survey. Additional essential tasks that could be completed before seeing the next patient during probe disinfection were also documented. Using another QR code-linked survey, data on ease of use, satisfaction with the system used and preferred system were collected. The ease of use and satisfaction with the system were rated on a 0 to 10 Likert scale (0 poor, 10 excellent). A free-text section for comments was then completed.

Results

Disinfection using UV-C (n = 331) was 60% faster than the chlorine dioxide multiwipe system (n = 332) (101 vs 250 s; P < 0.0001). A greater number of tasks were completed during probe disinfection when using UV-C, saving a further 74 s per patient (P < 0.0001). The HCPs using UV-C (n = 71) reported greater ease of use (median Likert score, 10 vs 3; P < 0.0001) and satisfaction (median Likert score, 10 vs 2; P < 0.0001) compared with those using the multiwipe system (n = 43). HCPs reported that the chlorine dioxide system was time-consuming and environmentally unfriendly, while the UV-C system was efficient and easy to use. Overall, 98% of the HCPs preferred using the UV-C system.

Conclusions

UV-C technology is more time-efficient and allows more essential tasks to be completed during disinfection. For a 4-h ultrasound list of 15 patients, the use of UV-C would save 55 min 45 s. HCPs found UV-C preferable and easier to use. © 2021 International Society of Ultrasound in Obstetrics and Gynecology.

SUBMITTER: Kyriacou C 

PROVIDER: S-EPMC9414347 | biostudies-literature | 2022 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Time-effectiveness and convenience of transvaginal ultrasound probe disinfection using ultraviolet vs chlorine dioxide multistep wipe system: prospective survey study.

Kyriacou C C   Robinson E E   Barcroft J J   Parker N N   Tuomey M M   Stalder C C   Gould D D   Al-Memar M M   Bourne T T  

Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology 20220701 1


<h4>Objectives</h4>To compare the efficiency, ease of use and user satisfaction of two methods of transvaginal ultrasound probe high-level disinfection: ultraviolet-C radiation (UV-C) and a chlorine dioxide multistep wipe system.<h4>Methods</h4>This was a prospective survey study. UV-C units were introduced into a busy early pregnancy assessment service and compared with a multiwipe system for disinfection. Before seeing each patient, healthcare professionals (HCPs) measured with a stopwatch the  ...[more]

Similar Datasets

| S-EPMC10890779 | biostudies-literature
| S-EPMC9770000 | biostudies-literature
| S-EPMC3947783 | biostudies-other
| S-EPMC5996494 | biostudies-literature
| S-EPMC8575156 | biostudies-literature
| S-EPMC4940628 | biostudies-literature
| S-EPMC3973690 | biostudies-other
2014-03-07 | GSE48187 | GEO
| S-EPMC9104377 | biostudies-literature
| S-EPMC7413072 | biostudies-literature