Accurate detection of heart rate using in-ear photoplethysmography in a clinical setting


ABSTRACT:

Background

Recent research has shown that photoplethysmography (PPG) based wearable sensors offer a promising potential for chronic disease monitoring. The aim of the present study was to assess the performance of an in-ear wearable PPG sensor in acquiring valid and reliable heart rate measurements in a clinical setting, with epileptic patients.

Methods

Patients undergoing video-electroencephalography (EEG) monitoring with concomitant one-lead electrocardiographic (ECG) recordings were equipped with an in-ear sensor developed by cosinuss°.

Results

In total, 2,048 h of recording from 97 patients with simultaneous ECG and in-ear heart rate data were included in the analysis. The comparison of the quality-filtered in-ear heart rate data with the reference ECG resulted in a bias of 0.78 bpm with a standard deviation of ±2.54 bpm; Pearson’s Correlation Coefficient PCC = 0.83; Intraclass Correlation Coefficient ICC = 0.81 and mean absolute percentage error MAPE = 2.57.

Conclusion

These data confirm that the in-ear wearable PPG sensor provides accurate heart rate measurements in comparison with ECG under realistic clinical conditions, especially with a signal quality indicator. Further research is required to investigate whether this technology is helpful in identifying seizure-related cardiovascular changes.

SUBMITTER: Adams T 

PROVIDER: S-EPMC9428405 | biostudies-literature | 2022 Jan

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC9604655 | biostudies-literature
| S-EPMC6118383 | biostudies-literature
| S-EPMC11902391 | biostudies-literature
| S-EPMC8160717 | biostudies-literature
| S-EPMC5820156 | biostudies-literature
| S-EPMC3949790 | biostudies-literature
| S-EPMC7177218 | biostudies-literature
| S-EPMC9267568 | biostudies-literature
| S-EPMC4228159 | biostudies-literature
| S-EPMC9370973 | biostudies-literature