Assessing a novel second generation laryngeal mask airway using the 'ADEPT' approach: results from the LMA® Protector™ observational study.
Ontology highlight
ABSTRACT: To address the problem of lack of clinical evidence for airway devices introduced to the market, the Difficult Airway Society (UK) developed an approach (termed ADEPT; Airway Device Evaluation Project Team) to standardise the model for device evaluation. Under this framework we assessed the LMA Protector, a second generation laryngeal mask airway. A total of 111 sequential adult patients were recruited and the LMA Protector inserted after induction of general anaesthesia. Effective insertion was confirmed by resistance to further distal movement, manual ventilation, and listening for gas leakage at the mouth. The breathing circuit was connected to the airway channel and airway patency confirmed with manual test ventilation at 20 cm H20 (water) pressure for 3 s. Data was collecte