A Modified CGM Algorithm Enhances Data Availability While Retaining iCGM Performance.
Ontology highlight
ABSTRACT: The algorithm for the Dexcom G6 CGM System was enhanced to retain accuracy while reducing the frequency and duration of sensor error. The new algorithm was evaluated by post-processing raw signals collected from G6 pivotal trials (NCT02880267) and by assessing the difference in data availability after a limited, real-world launch. Accuracy was comparable with the new algorithm-the overall %20/20 was 91.7% before and 91.8% after the algorithm modification; MARD was unchanged. The mean data gap due to sensor error nearly halved and total time spent in sensor error decreased by 59%. A limited field launch showed similar results, with a 43% decrease in total time spent in sensor error. Increased data availability may improve patient experience and CGM data integration into insulin delivery sys
SUBMITTER: Harvey Towers RA
PROVIDER: S-EPMC9445335 | biostudies-literature | 2022 Sep
REPOSITORIES: biostudies-literature
ACCESS DATA