Comparison of four different hematocrit assays and the effect of albumin on their measurements.


ABSTRACT: Clinical decisions are influenced by hematocrit values. Centrifugation (reference standard), conductivity, optical and impedance methods are often used interchangeably to measure hematocrit. The effects of albumin, which are known to affect conductivity methods, have not been evaluated for limits of agreement (LOA) between hematocrit assays in small animals. Canine venous blood was collected from 74 clinical cases and measured by centrifugation (n = 72), conductivity (n = 73), impedance (n = 24) and optical (n = 50) methods. Bland-Altman analysis determined bias (± SD) and 95% LOA between methods. There was a statistically significant difference between centrifugation hematocrit values and values obtained via conductivity (p < 0.0001), optical (