Real-time monitoring of isothermal nucleic acid amplification on a smartphone by using a portable electrochemical device for home-testing of SARS-CoV-2.
Ontology highlight
ABSTRACT: Home-testing of SARS-CoV-2 is an ideal approach for controlling the pandemic of COVID-19 and alleviating the shortage of medical resource caused by this acute infectious disease. Herein, a portable device that enables real-time monitoring of isothermal nucleic acid amplification tests (INAATs) through the electrochemistry method was fabricated for home-testing of SARS-CoV-2. First, a disposable plug-and-play pH-sensitive potentiometric sensor that matches this electrochemical INAATs (E-INAATs) device was designed to allow the label-free pH sensing detection of nucleic acid. By applying Nafion film on the polyaniline-based working electrode, this sensor exhibited an excellent linear potentiometric response to pH value in the range of 6.0-8.5 with a slope of -37.45 ± 1.96 mV/pH unit. A Bluet
SUBMITTER: Li Q
PROVIDER: S-EPMC9449873 | biostudies-literature | 2022 Oct
REPOSITORIES: biostudies-literature
ACCESS DATA