Unknown

Dataset Information

A field-portable colorimetric method for the measurement of peracetic acid vapors: a comparison of glass and plastic impingers.


ABSTRACT: A method for measuring peracetic acid vapors in air using impinger sampling and field-portable colorimetric analysis is presented. The capture efficiency of aqueous media in glass and plastic impingers was evaluated when used for peracetic acid vapor sampling. Measurement of peracetic acid was done using an N,N-diethyl-p-phenylenediamine colorimetric method with a field-portable spectrometer. The linearity of the N,N-diethyl-p-phenylenediamine method was determined for peracetic acid both in solution and captured from vapor phase using glass or plastic impingers. The Limits of Detection for the glass and plastic impingers were 0.24 mg/m3 and 0.28 mg/m3, respectively, for a 15 L air sample. The Limits of Quantitation were 0.79 mg/m3 and 0.92 mg

SUBMITTER: Stastny AL 

PROVIDER: S-EPMC9458618 | biostudies-literature | 2022 Aug

REPOSITORIES: biostudies-literature