A multicenter performance evaluation of the new Elecsys Vitamin D total III assay versus reference isotope dilution liquid chromatography tandem mass spectrometry and commercially available comparators.


ABSTRACT:

Background

Vitamin D deficiency/insufficiency and toxicity are worldwide issues; thus, accurate diagnostic assays are required to measure vitamin D. We evaluated the performance of the new Elecsys® Vitamin D total III assay (Roche Diagnostics International Ltd).

Methods

Repeatability and intermediate precision of the Elecsys Vitamin D total III assay (cobas e 601 analyzer) were evaluated at three sites using five human serum pools (HSPs) and two PreciControls (five-day model, one reagent lot [CLSI-EP05-A3]) and compared against prespecified acceptance criteria. A serum verification panel, with reference isotope dilution liquid chromatography tandem mass spectrometry (ID-LC-MS/MS) values, was used for comparator assay/concordance studies at two sites, assessed using unweighted Deming regression. Testing of serum vs. plasma on the Elecsys assay was conducted at one site using samples from healthy adults; assessed using Passing-Bablok regression.

Results

Repeatability (HSP1 [16.8-18.4 ng/ml], SD 0.87-1.07; HSP5 [94.5-98.0 ng/ml], CV 1.58%-2.76%) and intermediate precision (HSP1, SD 1.14-1.77; HSP5, CV 2.00%-4.13%) met acceptance criteria across sites. Agreement was observed between the Elecsys assay and (i) the ID-LC-MS/MS verification panel (slope, 0.936-1.01; Pearson's r, 0.960-0.986) and (ii) comparator assays (slope, 0.921-1.15; Pearson's r, 0.958-0.982). The Elecsys assay correctly assigned the highest combined percentage of samples to deficient (100%) and insufficient (89.5%) vitamin D categories vs. comparator assays and demonstrated comparable performance in serum and plasma (y = 0.103 + 0.984x).

Conclusions

The Elecsys Vitamin D total III assay demonstrated good analytical performance and compared favorably with other assays, supporting its use in clinical practice.

SUBMITTER: Findeisen P 

PROVIDER: S-EPMC9459274 | biostudies-literature | 2022 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

A multicenter performance evaluation of the new Elecsys Vitamin D total III assay versus reference isotope dilution liquid chromatography tandem mass spectrometry and commercially available comparators.

Findeisen Peter P   Leis Marta M   Bendig Garnet G   Grimme Jill J   Moser Elizabeth E   Vogl Christian C   Christenson Robert R  

Journal of clinical laboratory analysis 20220719 9


<h4>Background</h4>Vitamin D deficiency/insufficiency and toxicity are worldwide issues; thus, accurate diagnostic assays are required to measure vitamin D. We evaluated the performance of the new Elecsys<sup>®</sup> Vitamin D total III assay (Roche Diagnostics International Ltd).<h4>Methods</h4>Repeatability and intermediate precision of the Elecsys Vitamin D total III assay (cobas e 601 analyzer) were evaluated at three sites using five human serum pools (HSPs) and two PreciControls (five-day  ...[more]

Similar Datasets

| S-EPMC2652408 | biostudies-literature
| S-EPMC7180471 | biostudies-literature
| S-EPMC11377629 | biostudies-literature
| S-EPMC7339657 | biostudies-literature
| S-EPMC11294408 | biostudies-literature
| S-EPMC4823009 | biostudies-literature
| S-EPMC3727383 | biostudies-literature
| S-EPMC11609715 | biostudies-literature
| S-EPMC5362735 | biostudies-literature
| S-EPMC8190791 | biostudies-literature