A multicenter performance evaluation of the new Elecsys Vitamin D total III assay versus reference isotope dilution liquid chromatography tandem mass spectrometry and commercially available comparators.


ABSTRACT:

Background

Vitamin D deficiency/insufficiency and toxicity are worldwide issues; thus, accurate diagnostic assays are required to measure vitamin D. We evaluated the performance of the new Elecsys® Vitamin D total III assay (Roche Diagnostics International Ltd).

Methods

Repeatability and intermediate precision of the Elecsys Vitamin D total III assay (cobas e 601 analyzer) were evaluated at three sites using five human serum pools (HSPs) and two PreciControls (five-day model, one reagent lot [CLSI-EP05-A3]) and compared against prespecified acceptance criteria. A serum verification panel, with reference isotope dilution liquid chromatography tandem mass spectrometry (ID-LC-MS/MS) values, was used for comparator assay/concordance studies at two sites, assessed usin