Unknown

Dataset Information

Comparison of Prospectively Generated Glioma Treatment Plans Clinically Delivered on Magnetic Resonance Imaging (MRI)-Linear Accelerator (MR-Linac) Versus Conventional Linac: Predicted and Measured Skin Dose.


ABSTRACT: Introduction: Magnetic resonance imaging-linear accelerator radiotherapy is an innovative technology that requires special consideration for secondary electron interactions within the magnetic field, which can alter dose deposition at air-tissue interfaces. As part of ongoing quality assurance and quality improvement of new radiotherapy technologies, the purpose of this study was to evaluate skin dose modelled from the treatment planning systems of a magnetic resonance imaging-linear accelerator and a conventional linear accelerator, and then correlate with in vivo measurements of delivered skin dose from each linear accelerator. Methods: In this prospective cohort study, 37 consecutive glioma patients had treatment planning completed and approved prior to radiotherapy initia

SUBMITTER: Wang MH 

PROVIDER: S-EPMC9459463 | biostudies-literature | 2022 Jan-Dec

REPOSITORIES: biostudies-literature