Simplification of 2D shear wave elastography by enlarged SWE box and multiple regions of interest in one acquisition.


ABSTRACT: Two-dimensional (2D) Shear Wave Elastography (SWE) is an easy to perform technique to evaluate the liver stiffness. To simplify the procedure and reduce the acquisition time we enlarged the size of the SWE-box and set ten regions of interest (ROI) in one acquisition. We compare the accuracy of this method to ten separate acquisitions in a small box each with a single ROI measurement. Sixty-nine volunteers with diffuse chronic liver disease were studied with 2D-SWE using a Canon Aplio i800 ultrasound system. The shear-wave-speed was measured in the right lobe in ten separate acquisitions and compared to one acquisition with increased size of the SWE-box and ten different ROI measurements. A Bland-Altmann plot was drawn and the interclass correlation coefficient (ICC) was calculated to compare both methods. Finally, 2D-SWE was successfully performed thru both methods in sixty-six participants. Between both methods the ICC is 0.82. The results of this study show a good reliability between ten separate measures and one grouped measure with ten ROI if the mean is below 1.6m/s (7.7kPa). For higher degrees of fibrosis (≥F2) further investigations are needed.

SUBMITTER: Weiher M 

PROVIDER: S-EPMC9462759 | biostudies-literature | 2022

REPOSITORIES: biostudies-literature

altmetric image

Publications

Simplification of 2D shear wave elastography by enlarged SWE box and multiple regions of interest in one acquisition.

Weiher Moritz M   Richtering Felix Gunnar FG   Dörffel Yvonne Y   Müller Hans-Peter HP  

PloS one 20220909 9


Two-dimensional (2D) Shear Wave Elastography (SWE) is an easy to perform technique to evaluate the liver stiffness. To simplify the procedure and reduce the acquisition time we enlarged the size of the SWE-box and set ten regions of interest (ROI) in one acquisition. We compare the accuracy of this method to ten separate acquisitions in a small box each with a single ROI measurement. Sixty-nine volunteers with diffuse chronic liver disease were studied with 2D-SWE using a Canon Aplio i800 ultras  ...[more]

Similar Datasets

| S-EPMC7062958 | biostudies-literature
| S-EPMC10521701 | biostudies-literature
| S-EPMC4907490 | biostudies-literature
| S-EPMC8531167 | biostudies-literature
| S-EPMC10857377 | biostudies-literature
| S-EPMC10949614 | biostudies-literature
| S-EPMC7103049 | biostudies-literature
| S-EPMC8871512 | biostudies-literature
| S-EPMC7190848 | biostudies-literature
| S-EPMC5269522 | biostudies-literature