Dosimetric comparison between VMAT plans using the fast-rotating O-ring linac with dual-layer stacked MLC and helical tomotherapy for nasopharyngeal carcinoma.


ABSTRACT:

Background

To compare the dosimetric profiles of volumetric modulated arc therapy (VMAT) plans using the fast-rotating O-ring linac (the Halcyon system) based on a dual-layer stacked multi-leaf collimator and helical tomotherapy (HT) for nasopharyngeal cancer (NPCa).

Methods

For 30 NPCa patients, three sets of RT plans were generated, under the same policy of contouring and dose constraints: HT plan; Halcyon VMAT plan with two arcs (HL2arc); and Halcyon VMAT plan with four arcs (HL4arc), respectively. The intended dose schedule was to deliver 67.2 Gy to the planning gross target volume (P-GTV) and 56.0 Gy to the planning clinical target volume (P-CTV) in 28 fractions using the simultaneously integrated boost concept. Target volumes and organ at risks do