Unknown

Dataset Information

Hemodynamic numerical simulation of aortic arch modular inner branched stent-graft in eight early patients from the first-in-human case series.


ABSTRACT:

Background

The modular inner branched stent-graft (MIBSG) (WeFlow-Arch™) is an emerging device for challenging aortic arch pathologies. Hemodynamic numerical simulation is conducive to predicting long-term outcomes as well as optimizing the stent-graft design.

Objective

This study aims to analyze the hemodynamic characteristics of the MIBSG devices based on numerical simulation analyses.

Methods

From June 2019 to June 2021, MIBSGs were utilized in eight cases. Numerical simulation analyses of branch perfusion and indicators including the time-averaged wall shear stress, oscillatory shear index, and relative residence time were performed.

Results

Lesions involved Zone 1 (n = 2), Zone 2 (n = 4), and Zone 3 (n = 2). Branched stent-grafts were

SUBMITTER: Zhu Y 

PROVIDER: S-EPMC9468476 | biostudies-literature | 2022