ABSTRACT: Purpose
This study will evaluate the clinical quality and usability of peripheral image data from the temporal bone area obtained using a sinonasal ultra-low-dose (ULD) cone-beam computed tomography (CBCT) scan and compare them to those obtained using a high-resolution (HR) CBCT.Methods
The population consisted of 66 anatomical sites (ears of 33 subjects) imaged using two modalities: an HR CBCT (Scanora 3Dx scanner; Soredex, Tuusula, Finland) and a ULD CBCT (Promax 3D Mid scanner; Plandent, Helsinki, Finland). The image quality (IQ) for every anatomical site in each image was rated using a Likert scale from 0 to 5.Results
The quality of ULD CBCT scans was clinically sufficient in over 95% of the assessed images of the sigmoid sinus, jugular bulb, epitympanum and mas