Flexible endoscopy in the visualization of 3D-printed maxillary sinus and clinical application.


ABSTRACT:

Background

During postoperative follow-up, the visible range of maxillary sinus (MS) is limited, even combining 0° and 70° rigid endoscopes together. Flexible endoscope has been used in larynx examinations for a long time, but rarely in nasal cavity and sinus. We aimed to evaluate the application values of rigid and flexible endoscopes for visualization of MS.

Methods

We followed up 70 patients with lesions in MS via both rigid and flexible endoscopes. In addition, we used thin-slice CT image of the sinus to create a MS model and divided it into two parts for 3D printing. The inner surface of the 3D-printed sinus was marked with grid papers of the same size (5 mm × 5 mm), then the visual range under rigid endoscopes with different angle and flexible endoscopes was calculated