Unknown

Dataset Information

Feasibility of Blood Pressure Measurement With a Wearable (Watch-Type) Monitor During Impending Syncopal Episodes.


ABSTRACT: Background We assessed the reliability and feasibility of blood pressure (BP) measurements by means of a new wearable watch-type BP monitor (HeartGuide) in detecting episodes of hypotensive (pre)syncope induced by tilt table test. Methods and Results An intrapatient comparison between systolic BP (SBP) measured by means of the HeartGuide device and noninvasive finger beat-to-beat BP monitoring was undertaken both at baseline in supine position and repeatedly during tilt table test in patients evaluated for reflex syncope. Intrapatient fall of systolic BP from baseline was measured. Eighty-one patients (mean age, 61±19 years; 46 women) were included. Overall, HeartGuide was able to yield BP values at the time of BP nadir in 58 (72%) patients (average HeartGuide SBP 102±18 mm Hg, versus fing

SUBMITTER: Groppelli A 

PROVIDER: S-EPMC9496316 | biostudies-literature | 2022 Aug

REPOSITORIES: biostudies-literature