ABSTRACT: Purpose
Multi-leaf-collimator (MLC) leaf position accuracy is important for accurate dynamic radiotherapy treatment plan delivery. Machine log files have become widely utilized for quality assurance (QA) of such dynamic treatments. The primary aim is to test the sensitivity of machine log files in comparison to electronic portal imaging device (EPID)-based measurements to MLC position errors caused by leaf backlash. The secondary aim is to investigate the effect of MLC leaf backlash on MLC leaf motion during clinical dynamic plan delivery.Methods
The sensitivity of machine log files and two EPID-based measurements were assessed via a controlled experiment, whereby the length of the "T" section of a series of 12 MLC leaf T-nuts in a Varian Millennium MLC for a Trilogy C-seri