Does augmentation increase the pull-out force of symphyseal screws? A biomechanical cadaver study.


ABSTRACT:

Purpose

Open reduction and internal fixation using anterior plate osteosynthesis currently represents the gold standard for the treatment of symphyseal disruptions. Since postoperative screw loosening with consequent implant failure is frequently observed, this study aims to evaluate if and to what extent augmentation can increase the pull-out force of symphyseal screws to improve the constructs stability.

Methods

Twelve human cadaveric anterior pelvic rings were separated at the symphyseal joint for bilateral testing, consequently achieving comparable sites. First, one non-augmented screw was drilled into the superior pubic ramus, whereas the contralateral side was primarily augmented. The screws were then withdrawn with a constant speed of 10 mm/min and the fixation streng

SUBMITTER: Cavalcanti Kußmaul A 

PROVIDER: S-EPMC9532288 | biostudies-literature | 2022 Oct

REPOSITORIES: biostudies-literature