Intra-beat biomarker for accurate continuous non-invasive blood pressure monitoring.
Ontology highlight
ABSTRACT: Accurate continuous non-invasive blood pressure (CNIBP) monitoring is the holy grail of digital medicine but remains elusive largely due to significant drifts in signal and motion artifacts that necessitate frequent device recalibration. To address these challenges, we developed a unique approach by creating a novel intra-beat biomarker (Diastolic Transit Time, DTT) to achieve highly accurate blood pressure (BP) estimations. We demonstrated our approach's superior performance, compared to other common signal processing techniques, in eliminating stochastic baseline wander, while maintaining signal integrity and measurement accuracy, even during significant hemodynamic changes. We applied this new algorithm to BP data collected using non-invasive sensors from a diverse cohort of high acuity
SUBMITTER: Abiri A
PROVIDER: S-EPMC9537243 | biostudies-literature | 2022 Oct
REPOSITORIES: biostudies-literature