Efficacy of a novel channel-cleaning ball brush for endoscope reprocessing: a randomized controlled trial.


ABSTRACT:

Background/aims

Endoscopic channels are difficult to clean and can cause infection transmission. We examined the effectiveness of a newly developed channel-cleaning ball brush (BB), which is sucked into the endoscopic channel and scrapes and cleans the lumen as it passes through.

Methods

The upper and lower gastrointestinal endoscopes used for patient examinations were randomly selected as the conventional brush (CB) or BB group. After manual cleaning, the presence or absence of carbohydrates, proteins, adenosine triphosphate, and hemoglobin was assessed.

Results

Fifty-six and 58 endoscopes were cleaned with the CB and BB, respectively. Carbohydrate and protein were detected in one (1.8%) and two endoscopes (3.4%) in the CB and BB groups, respectively (p=1.000). Hemoglobin was observed in one (1.8%) and three endoscopes (5.2%) in the CB and BB groups, respectively (p=0.636). The adenosine triphosphate levels were 10.6±15.9 and 12.5±14.3 relative light units in the CB and BB groups, respectively (p=0.496). Twenty-seven (48.2%) and 19 (32.8%) endoscopes were positive for microbial cultures in the CB and BB groups, respectively (p=0.136).

Conclusion

The efficacy of BB was not significantly different from that of CB in the endoscopic channel-cleaning process.

SUBMITTER: Chung KH 

PROVIDER: S-EPMC9539290 | biostudies-literature | 2022 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Efficacy of a novel channel-cleaning ball brush for endoscope reprocessing: a randomized controlled trial.

Chung Kwang Hyun KH   Chae Jeong Don JD   Choe Wonho W   Lee Hyo Young HY   Oh Il Hwan IH   Son Byoung Kwan BK  

Clinical endoscopy 20220802 5


<h4>Background/aims</h4>Endoscopic channels are difficult to clean and can cause infection transmission. We examined the effectiveness of a newly developed channel-cleaning ball brush (BB), which is sucked into the endoscopic channel and scrapes and cleans the lumen as it passes through.<h4>Methods</h4>The upper and lower gastrointestinal endoscopes used for patient examinations were randomly selected as the conventional brush (CB) or BB group. After manual cleaning, the presence or absence of c  ...[more]

Similar Datasets

| S-EPMC11583001 | biostudies-literature
| S-EPMC11837676 | biostudies-literature
| S-EPMC6296091 | biostudies-literature
| S-EPMC6748329 | biostudies-literature
| S-EPMC5533832 | biostudies-other
| S-EPMC6363334 | biostudies-literature
2024-06-13 | PXD046061 | Pride
| S-EPMC8668382 | biostudies-literature
| S-EPMC10507511 | biostudies-literature
| S-EPMC11922170 | biostudies-literature