Subject-specific optimization of background suppression for arterial spin labeling magnetic resonance imaging using a feedback loop on the scanner.
Ontology highlight
ABSTRACT: Background suppression (BGS) in arterial spin labeling (ASL) magnetic resonance imaging leads to a higher temporal signal-to-noise ratio (tSNR) of the perfusion images compared with ASL without BGS. The performance of the BGS, however, depends on the tissue relaxation times and on inhomogeneities of the scanner's magnetic fields, which differ between subjects and are unknown at the moment of scanning. Therefore, we developed a feedback loop (FBL) mechanism that optimizes the BGS for each subject in the scanner during acquisition. We implemented the FBL for 2D pseudo-continuous ASL scans with an echo-planar imaging readout. After each dynamic scan, the acquired ASL images were automatically sent to an external computer and processed with a Python processing tool. Inversion times were optimi