Cortical auditory evoked potential in cochlear implant users: An objective method to improve speech perception.


ABSTRACT:

Objective

To investigate if cortical auditory evoked potential (CAEP) measures can be used to verify the cochlear implant (CI) map and consequently improve CI outcomes in adults with bilateral hearing loss.

Design

CAEPs were measured in CI recipients using the speech tokens /m/, /g/, /t/ and /s/. If CAEP responses were present for all speech tokens, the participant's map was considered "satisfactory". If CAEP responses were absent, the CI map was considered "unsatisfactory" and therefore adjusted and CAEP measures repeated. This was repeated until auditory potentials were seen in response to all four speech tokens. Speech testing was conducted pre-CI, as well as before and after CAEP-guided map adjustments.

Results

108 adult unilateral CI users participated, whose sou