Ontology highlight
ABSTRACT: Purpose
To characterize the acceleration capabilities of a silent head insert gradient axis that operates at the inaudible frequency of 20 kHz and a maximum gradient amplitude of 40 mT/m without inducing peripheral nerve stimulation.Methods
The silent gradient axis' acquisitions feature an oscillating gradient in the phase-encoding direction that is played out on top of a cartesian readout, similarly as done in Wave-CAIPI. The additional spatial encoding fills k-space in readout lanes allowing for the acquisition of fewer phase-encoding steps without increasing aliasing artifacts. Fully sampled 2D gradient echo datasets were acquired both with and without the silent readout. All scans were retrospectively undersampled (acceleration factors R = 1 to 12) to compare convention
SUBMITTER: Versteeg E
PROVIDER: S-EPMC9544176 | biostudies-literature | 2022 Oct
REPOSITORIES: biostudies-literature