Comparison of MiniMed 780G system performance in users aged younger and older than 15 years: Evidence from 12 870 real-world users.


ABSTRACT:

Aim

To investigate real-world glycaemic outcomes and goals achieved by users of the MiniMed 780G advanced hybrid closed loop (AHCL) system aged younger and older than 15 years with type 1 diabetes (T1D).

Materials and methods

Data uploaded by MiniMed 780G system users from 27 August 2020 to 22 July 2021 were aggregated and retrospectively analysed based on self-reported age (≤15 years and >15 years) for three cohorts: (a) post-AHCL initiation, (b) 6-month longitudinal post-AHCL initiation and (c) pre- versus post-AHCL initiation. Analyses included mean percentage of time spent in AHCL and at sensor glucose ranges, insulin delivered and the proportion of users achieving recommended glucose management indicator (GMI < 7.0%) and time in target range (TIR 70-180 mg/dl > 70%) goa