The diagnostic accuracy of Sysmex XN for identification of pseudothrombocytopenia using various thresholds for definition of platelet aggregation.


ABSTRACT:

Introduction

The aim of this study was to investigate the diagnostic accuracy of the flag PLT-Clumps from the WNR/WDF and the PLT-F channels from Sysmex XN and to study how different cut-offs for investigation for pseudothrombocytopenia (PTCP) and the definition of platelet aggregation affected the diagnostic accuracy.

Methods

A smear review was performed for samples with platelet count <150 × 109 /L and samples flagged for platelet aggregation by Sysmex XN-20. The samples were investigated for platelet aggregation in 30 fields using 40× objective. Findings were classified by size and quantity using two definitions of aggregation: the Norwegian quality improvement of laboratory investigations (Noklus) and the Groupe Francophone de'Hematologie Cellulaire (GFHC). Th