ABSTRACT: Background
Insertable cardiac monitors (ICMs) are small subcutaneously implanted devices that detect changes in R-wave amplitudes (RWAs), effective in arrhythmia-monitoring. Although ICMs have proven to be immensely successful, electrical artefacts are frequent and can lead to misdiagnosis. Thus, there is a growing need to sustain and increase efficacy in detection rates by gaining insight into various patient-specific factors such as body postures and activities.Methods
RWAs were measured in 15 separate postures, including supine, lying on the right-side (RS) or left-side (LS) and sitting, and two separate ICM orientations, immediately after implantation of Confirm Rx™ ICM in 99 patients.Results
The patients (53 females and 46 males, mean ages 66.62 ± 14.7 and 66.4