Deep Single-Shot NanoLC-MS Proteome Profiling with a 1500 Bar UHPLC System, Long Fully Porous Columns, and HRAM MS.


ABSTRACT: This study demonstrates how the latest ultrahigh-performance liquid chromatography (UHPLC) technology can be combined with high-resolution accurate-mass (HRAM) mass spectrometry (MS) and long columns packed with fully porous particles to improve bottom-up proteomics analysis with nanoflow liquid chromatography-mass spectrometry (nanoLC-MS) methods. The increased back pressures from the UHPLC system enabled the use of 75 μm I.D. × 75 cm columns packed with 2 μm particles at a typical 300 nL/min flow rate as well as elevated and reduced flow rates. The constant pressure pump operation at 1500 bar reduced sample loading and column washing/equilibration stages and overall overhead time, which maximizes MS utilization time. The versatility of flow rate optimization to balance the sensitivity, t